Safety Alert for Patient monitor Procare

According to National Institute of Drugs and Food Surveillance (INVIMA), this safety alert involved a device in Colombia that was produced by GE Medical Systems Technologies Co Ltd || GE Healthcare Finland OY.

What is this?

Alerts provide important information and recommendations about products. Even though an alert has been issued, it does not necessarily mean a product is considered to be unsafe. Safety Alerts, addressed to health workers and users, may include recalls. They can be written by manufacturers, but also by health officials.

Learn more about the data here
  • Type of Event
    Safety alert
  • Event ID
    I1402-85
  • Event Number
    INVIMA 2008EBC-0001978
  • Date
    2014-02-24
  • Event Country
  • Event Source
    INVIMA
  • Event Source URL
  • Notes / Alerts
    Colombian data is current through June 2019. All of the data comes from the National Institute of Drugs and Food Surveillance (INVIMA), except for the categories Manufacturer Parent Company and Product Classification.
    The Parent Company and the Product Classification were added by ICIJ.
    The parent company information is based on 2017 public records. The device classification information comes from FDA’s Product Classification by Review Panel, based on matches of recall data from the U.S. and Colombia.
  • Extra notes in the data
  • Reason
    The manufacturer informs that the filters for the ecg signal selected by the users can be affected by an additional high-pass filter of 0.5 hz, which would generate changes in the ecg wave morphology and values on the screen of stincorrect, leading to there are potentially adverse events on the patient.

Device

Manufacturer