Safety Alert for Mitraclip System

According to National Institute of Drugs and Food Surveillance (INVIMA), this safety alert involved a device in Colombia that was produced by Evalve Inc..

What is this?

Alerts provide important information and recommendations about products. Even though an alert has been issued, it does not necessarily mean a product is considered to be unsafe. Safety Alerts, addressed to health workers and users, may include recalls. They can be written by manufacturers, but also by health officials.

Learn more about the data here
  • Type of Event
    Safety alert
  • Event ID
    I1602-50
  • Event Number
    2012DM-0009235
  • Date
    2016-02-11
  • Event Country
  • Event Source
    INVIMA
  • Event Source URL
  • Notes / Alerts
    Colombian data is current through June 2019. All of the data comes from the National Institute of Drugs and Food Surveillance (INVIMA), except for the categories Manufacturer Parent Company and Product Classification.
    The Parent Company and the Product Classification were added by ICIJ.
    The parent company information is based on 2017 public records. The device classification information comes from FDA’s Product Classification by Review Panel, based on matches of recall data from the U.S. and Colombia.
  • Extra notes in the data
  • Reason
    The manufacturer states that in the referenced devices a mandrel fracture could occur when the tension is present in the mandrel when turning the actuator knob to unfold the clip, the tension is present if the arm positioner is "closed" in the neutral position, instead to be in the neutral position during the deployment of the clip, leading to potential adverse events being presented to patients.

Device

  • Model / Serial
    CDS02ST and SGC01ST, lots greater than 50714U1
  • Product Classification
  • Product Description
    Implantation system for the reconstruction of insufficient mitral valves by tissue approximation.
  • Manufacturer

Manufacturer

  • Manufacturer Parent Company (2017)
  • Manufacturer comment
    “We are in constant communication with regulatory agencies and competent authorities worldwide which allows us to implement global recalls or in-country communication quickly and effectively,” Abbott, which now owns St. Jude Medical told ICIJ in a statement. In addition to sending global notices to physicians worldwide, we also make sure that product advisories are available online and classification of product recalls and product advisories are determined by global regulatory bodies which can impact the timing in any given country. MD companies follow varying regulations in different countries. In come countries software is not regulated so a recall in one country related to software would not be classified as a recall or field action in another. In addition, review cycles within the regulatory process can be different in each country which can impact communication and recall timing.
  • Source
    NIDFSINVIMA