Safety Alert for MICROLINE SURGICAL RE-NEW Tips

According to National Institute of Drugs and Food Surveillance (INVIMA), this safety alert involved a device in Colombia that was produced by Microline Surgical, Inc..

What is this?

Alerts provide important information and recommendations about products. Even though an alert has been issued, it does not necessarily mean a product is considered to be unsafe. Safety Alerts, addressed to health workers and users, may include recalls. They can be written by manufacturers, but also by health officials.

Learn more about the data here
  • Type of Event
    Safety alert
  • Event ID
    R1708-355
  • Event Number
    2010DM-0006298
  • Date
    2017-10-24
  • Event Country
  • Event Source
    INVIMA
  • Event Source URL
  • Notes / Alerts
    Colombian data is current through June 2019. All of the data comes from the National Institute of Drugs and Food Surveillance (INVIMA), except for the categories Manufacturer Parent Company and Product Classification.
    The Parent Company and the Product Classification were added by ICIJ.
    The parent company information is based on 2017 public records. The device classification information comes from FDA’s Product Classification by Review Panel, based on matches of recall data from the U.S. and Colombia.
  • Extra notes in the data
  • Reason
    The manufacturer indicates that due to the malfunction related to the material located in the proximal area of the metal lid, the product has been withdrawn, since after being subjected to heat in the sterilization process, it can contract and cause its rupture, leading to its occurrence possible adverse events on the patients.

Device

  • Model / Serial
    3231, 3241, 3251, 3361, 3341, 3411, 3431, all batches.
  • Product Description
    The tips and the RENEW handpiece are indicated for cutting, grasping, dissecting and coagulating tissues in endoscopic and laparoscopic surgical procedures.
  • Manufacturer

Manufacturer

  • Manufacturer Parent Company (2017)
  • Source
    NIDFSINVIMA