Safety Alert for Linear Accelerator PRIMUS / ONCOR SIEMENS

According to National Institute of Drugs and Food Surveillance (INVIMA), this safety alert involved a device in Colombia that was produced by Siemens Medical Solutions Usa, Inc..

What is this?

Alerts provide important information and recommendations about products. Even though an alert has been issued, it does not necessarily mean a product is considered to be unsafe. Safety Alerts, addressed to health workers and users, may include recalls. They can be written by manufacturers, but also by health officials.

Learn more about the data here
  • Type of Event
    Safety alert
  • Event ID
    A1501-42
  • Event Number
    2008EBC-0001814
  • Date
    2015-01-22
  • Event Country
  • Event Source
    INVIMA
  • Event Source URL
  • Notes / Alerts
    Colombian data is current through June 2019. All of the data comes from the National Institute of Drugs and Food Surveillance (INVIMA), except for the categories Manufacturer Parent Company and Product Classification.
    The Parent Company and the Product Classification were added by ICIJ.
    The parent company information is based on 2017 public records. The device classification information comes from FDA’s Product Classification by Review Panel, based on matches of recall data from the U.S. and Colombia.
  • Extra notes in the data
  • Reason
    The manufacturer has identified the possible occurrence of sporadic failures due to the overheating of an electronic component (pcb controller head), belonging to the printed circuit board on the main card g41, located inside the housing of the equipment, which may lead to the occurrence of potentially adverse events. about the patient or the user.

Device

  • Model / Serial
    PRIMUS, ONCOR and ARTISTE.
  • Product Description
    Ionizing radiation emission equipment for therapeutic treatment of cancer.
  • Manufacturer

Manufacturer

  • Manufacturer Parent Company (2017)
  • Source
    NIDFSINVIMA