Safety Alert for Equipment PACS

According to National Institute of Drugs and Food Surveillance (INVIMA), this safety alert involved a device in Colombia that was produced by Siemens AG || Siemens Medical Solutions Usa, Inc..

What is this?

Alerts provide important information and recommendations about products. Even though an alert has been issued, it does not necessarily mean a product is considered to be unsafe. Safety Alerts, addressed to health workers and users, may include recalls. They can be written by manufacturers, but also by health officials.

Learn more about the data here
  • Type of Event
    Safety alert
  • Event ID
    I1502-60
  • Event Number
    2010DM-0005218
  • Date
    2015-02-09
  • Event Country
  • Event Source
    INVIMA
  • Event Source URL
  • Notes / Alerts
    Colombian data is current through June 2019. All of the data comes from the National Institute of Drugs and Food Surveillance (INVIMA), except for the categories Manufacturer Parent Company and Product Classification.
    The Parent Company and the Product Classification were added by ICIJ.
    The parent company information is based on 2017 public records. The device classification information comes from FDA’s Product Classification by Review Panel, based on matches of recall data from the U.S. and Colombia.
  • Extra notes in the data
  • Reason
    The manufacturer states that in case of failure of the system (for example, blue screen or power cut), the images are not recorded on the hard disk, which potentially results in inconsistencies in the database causing its loss, leading to the occurrence possible adverse events on patients.

Device

Manufacturer

  • Manufacturer Parent Company (2017)
  • Source
    NIDFSINVIMA