Safety Alert for Engines for Surgical Use

According to National Institute of Drugs and Food Surveillance (INVIMA), this safety alert involved a device in Colombia that was produced by Synthes Gmbh.

What is this?

Alerts provide important information and recommendations about products. Even though an alert has been issued, it does not necessarily mean a product is considered to be unsafe. Safety Alerts, addressed to health workers and users, may include recalls. They can be written by manufacturers, but also by health officials.

Learn more about the data here
  • Type of Event
    Safety alert
  • Event ID
    R1602-78
  • Event Number
    2008DM-0003014
  • Date
    2016-02-24
  • Event Country
  • Event Source
    INVIMA
  • Event Source URL
  • Notes / Alerts
    Colombian data is current through June 2019. All of the data comes from the National Institute of Drugs and Food Surveillance (INVIMA), except for the categories Manufacturer Parent Company and Product Classification.
    The Parent Company and the Product Classification were added by ICIJ.
    The parent company information is based on 2017 public records. The device classification information comes from FDA’s Product Classification by Review Panel, based on matches of recall data from the U.S. and Colombia.
  • Extra notes in the data
  • Reason
    The manufacturer reports that it has detected that the charger has a component that can fail prematurely due to manufacturing failure, if the component fails, the firmware of the charger recognizes the malfunction and prevents the batteries from charging, the blue led on the front of the charger flashes to indicate that the device should be sent for repair, leading to potential adverse events on patients or delays in surgical times.

Device

  • Model / Serial
    UNIVERSAL BATTERY CHARGER II (UBC II), references 05.001.204, lots 10615 to 12616.
  • Product Description
    The engines for surgical use are tools that are intended to generate a continuous movement of their accessories that are cutting tools such as drills, bits and saws in orthopedic and traumatology procedures such as: placement and extraction of prosthesis and osteosynthesis material, intramedullary milling, traumatology in general, foot and hand surgery, surgery, maxillofacial neurosurgery and vertebral surgery
  • Manufacturer

Manufacturer