Safety Alert for emergency ventilator and transport

According to National Institute of Drugs and Food Surveillance (INVIMA), this safety alert involved a device in Colombia that was produced by Drägerwerk Ag & Co. Kgaa.

What is this?

Alerts provide important information and recommendations about products. Even though an alert has been issued, it does not necessarily mean a product is considered to be unsafe. Safety Alerts, addressed to health workers and users, may include recalls. They can be written by manufacturers, but also by health officials.

Learn more about the data here
  • Type of Event
    Safety alert
  • Event ID
    A1601-9
  • Event Number
    2009EBC-0003111 ; 2009EBC-0003135 ; 2011EBC-0006883
  • Date
    2016-01-07
  • Event Country
  • Event Source
    INVIMA
  • Event Source URL
  • Notes / Alerts
    Colombian data is current through June 2019. All of the data comes from the National Institute of Drugs and Food Surveillance (INVIMA), except for the categories Manufacturer Parent Company and Product Classification.
    The Parent Company and the Product Classification were added by ICIJ.
    The parent company information is based on 2017 public records. The device classification information comes from FDA’s Product Classification by Review Panel, based on matches of recall data from the U.S. and Colombia.
  • Extra notes in the data
  • Reason
    The manufacturer has detected that when using the equipment the message "poti unplugged" (connected potentiometer) can be generated, in these cases, an audible and visual alarm is generated, the pressure drops and the respiratory ventilation function stops, which can lead to their being displayed. potentially serious adverse events on the patients.

Device

  • Model / Serial
    OXYLOG 3000, OXYLOG 3000 PLUS and OXYLOG 2000 PLUS
  • Product Classification
  • Product Description
    Indicated for emergency ventilation and transportation cycled by time, controlled by volume and pressure support for patients who need mandatory or assisted ventilation.
  • Manufacturer

Manufacturer