Safety Alert for conventional mobile x-ray systems Practix 160

According to National Institute of Drugs and Food Surveillance (INVIMA), this safety alert involved a device in Colombia that was produced by Philips Medical Systems Dmc Gmbh.

What is this?

Alerts provide important information and recommendations about products. Even though an alert has been issued, it does not necessarily mean a product is considered to be unsafe. Safety Alerts, addressed to health workers and users, may include recalls. They can be written by manufacturers, but also by health officials.

Learn more about the data here
  • Type of Event
    Safety alert
  • Event ID
    I1503-110
  • Event Number
    2008EBC-0003048
  • Date
    2015-03-15
  • Event Country
  • Event Source
    INVIMA
  • Event Source URL
  • Notes / Alerts
    Colombian data is current through June 2019. All of the data comes from the National Institute of Drugs and Food Surveillance (INVIMA), except for the categories Manufacturer Parent Company and Product Classification.
    The Parent Company and the Product Classification were added by ICIJ.
    The parent company information is based on 2017 public records. The device classification information comes from FDA’s Product Classification by Review Panel, based on matches of recall data from the U.S. and Colombia.
  • Extra notes in the data
  • Reason
    The manufacturer states that when the previous systems are disconnected or when the arm of the tube is located in stand by (x-ray generator off), a pulse of radiation (50 kv, 2.5 μgy) can be generated, potentially resulting in the unintentional exposure, this occurs only in systems containing an electronic component in particular in the control board; the component is used only in a small number of systems, leading to the potential occurrence of adverse events on patients or users.

Device

  • Model / Serial
    PRACTIX 160
  • Product Description
    This equipment is indicated for the visualization and acquisition of X-ray images for diagnosis.
  • Manufacturer

Manufacturer

  • Manufacturer Parent Company (2017)
  • Source
    NIDFSINVIMA