What is this?

Alerts provide important information and recommendations about products. Even though an alert has been issued, it does not necessarily mean a product is considered to be unsafe. Safety Alerts, addressed to health workers and users, may include recalls. They can be written by manufacturers, but also by health officials.

Learn more about the data here
  • Type of Event
    Safety alert
  • Event ID
    I1609-390
  • Event Number
    2008EBC-0001540
  • Date
    2016-09-14
  • Event Country
  • Event Source
    INVIMA
  • Event Source URL
  • Notes / Alerts
    Colombian data is current through June 2019. All of the data comes from the National Institute of Drugs and Food Surveillance (INVIMA), except for the categories Manufacturer Parent Company and Product Classification.
    The Parent Company and the Product Classification were added by ICIJ.
    The parent company information is based on 2017 public records. The device classification information comes from FDA’s Product Classification by Review Panel, based on matches of recall data from the U.S. and Colombia.
  • Extra notes in the data
  • Reason
    The manufacturer affirms that if additional scans are necessary prior to completing the current reconstruction, noting the planned images of the current reconstruction are shown to the user, even though the information is not separated, the gross information can be reconstructed off-line showing all the reconstructions requested, however, users may be unaware that the raw information is still available for offline reconstruction, leading to new scans and possible adverse events on patients.

Device