Safety Alert for AISYS Anesthesia Machine

According to National Institute of Drugs and Food Surveillance (INVIMA), this safety alert involved a device in Colombia that was produced by Ge Medical Systems China Co, Ltd. || Datex - Ohmeda Inc..

What is this?

Alerts provide important information and recommendations about products. Even though an alert has been issued, it does not necessarily mean a product is considered to be unsafe. Safety Alerts, addressed to health workers and users, may include recalls. They can be written by manufacturers, but also by health officials.

Learn more about the data here
  • Type of Event
    Safety alert
  • Event ID
    I1604-144
  • Event Number
    2008EBC-0001418
  • Date
    2016-04-11
  • Event Country
  • Event Source
    INVIMA
  • Event Source URL
  • Notes / Alerts
    Colombian data is current through June 2019. All of the data comes from the National Institute of Drugs and Food Surveillance (INVIMA), except for the categories Manufacturer Parent Company and Product Classification.
    The Parent Company and the Product Classification were added by ICIJ.
    The parent company information is based on 2017 public records. The device classification information comes from FDA’s Product Classification by Review Panel, based on matches of recall data from the U.S. and Colombia.
  • Extra notes in the data
  • Reason
    The manufacturer has detected a problem in the suction system, where the tube is bent inside the equipment, generating a suction flow of less than 20 l / min, the value required by iso 10079-3, if this problem is not solved, the suction flow may not be sufficient to clear the airway and ventilate the patient properly, which could lead to potential delays in patient care.