Safety Alert for AGAR COLUMBIA BASE

According to National Institute of Drugs and Food Surveillance (INVIMA), this safety alert involved a device in Colombia that was produced by OXOID LIMITED || Imported by: QUIMIREL LTDA.

What is this?

Alerts provide important information and recommendations about products. Even though an alert has been issued, it does not necessarily mean a product is considered to be unsafe. Safety Alerts, addressed to health workers and users, may include recalls. They can be written by manufacturers, but also by health officials.

Learn more about the data here
  • Type of Event
    Safety alert
  • Event ID
    IRD-010916
  • Event Number
    INVIMA 2007RD-0000652
  • Date
    2016-09-01
  • Event Country
  • Event Source
    INVIMA
  • Event Source URL
  • Notes / Alerts
    Colombian data is current through June 2019. All of the data comes from the National Institute of Drugs and Food Surveillance (INVIMA), except for the categories Manufacturer Parent Company and Product Classification.
    The Parent Company and the Product Classification were added by ICIJ.
    The parent company information is based on 2017 public records. The device classification information comes from FDA’s Product Classification by Review Panel, based on matches of recall data from the U.S. and Colombia.
  • Extra notes in the data
  • Reason
    Customer complaint records and internal investigations have shown that the aforementioned batches of the columbia base agar cm0331 product (ref b, r, s, t, u) did not receive sufficient microbiological supplement, which results in colony recovery being less than expected also, the hemolysis is reduced or absent, therefore the use of these batches may prevent the identification or confirmation of the presence of certain bacterial strains of staphylococcus and streptococcus spp.

Device

Manufacturer