Recall of ZOLL AED PLUS DEFIBRILLATOR

According to Health Canada (via FOI), this recall involved a device in Canada that was produced by ZOLL MEDICAL CORPORATION.

What is this?

A correction or removal action taken by a manufacturer to address a problem with a medical device. Recalls occur when a medical device is defective, when it could be a risk to health, or when it is both defective and a risk to health.

Learn more about the data here
  • Type of Event
    Recall
  • Event ID
    28650
  • Event Risk Class
    II
  • Event Initiated Date
    2002-12-30
  • Event Country
  • Event Source
    HC
  • Notes / Alerts
    Canadian data is current through March 2018. All of the data comes from Health Canada, except for the categories Manufacturer Parent Company and Product Classification.
    The Parent Company and the Product Classification were added by ICIJ.
    The parent company information is based on 2017 public records. The device classification information comes from FDA’s Product Classification by Review Panel, based on matches of recall data from the U.S. and Canada.
  • Extra notes in the data
  • Reason
    Device analyses patientemits"shock advised" message. no charge is delivered and "no treatment delivered" or "change batteries" message is emitted.This occurs during electrode handling before they are attached.

Device

  • Model / Serial
    Model Catalog: (Lot serial: X02F000812 TO X02K007486)
  • Product Description
    ZOLL AED PLUS EXTERNAL DEFIBRILLATOR
  • Manufacturer

Manufacturer

  • Manufacturer Address
    CHELMSFORD
  • Manufacturer Parent Company (2017)
  • Source
    HC