Recall of XIMATRON

According to Health Canada (via FOI), this recall involved a device in Canada that was produced by VARIAN MEDICAL SYSTEMS INC..

What is this?

A correction or removal action taken by a manufacturer to address a problem with a medical device. Recalls occur when a medical device is defective, when it could be a risk to health, or when it is both defective and a risk to health.

Learn more about the data here
  • Type of Event
    Recall
  • Event ID
    54259
  • Event Risk Class
    II
  • Event Initiated Date
    2010-05-05
  • Event Country
  • Event Source
    HC
  • Notes / Alerts
    Canadian data is current through March 2018. All of the data comes from Health Canada, except for the categories Manufacturer Parent Company and Product Classification.
    The Parent Company and the Product Classification were added by ICIJ.
    The parent company information is based on 2017 public records. The device classification information comes from FDA’s Product Classification by Review Panel, based on matches of recall data from the U.S. and Canada.
  • Extra notes in the data
  • Reason
    The c-series ximatron uses a lead screw drive assembly system for the focus to axis movement and for the image intensifier vertical movement. under normal use these drive systems will wear and eventually require replacement. however failure to monitor the extent of the wear of the lead screw drive assemblies could lead to a failure of the lead screw drive system. this could potentially lead to a free floating drive system that could slide on the captive gantry rails during gantry rotation when operating one or both of these drive systems potentially leading to a collision with a patient an operator the floor or the treatment table.

Device

  • Model / Serial
    Model Catalog: CX (Lot serial: > 10 units. Contact); Model Catalog: CX (Lot serial: manufacturer.)
  • Product Description
    XIMATRON