Recall of WIRELESS XRAY FSW MONO

According to Health Canada (via FOI), this recall involved a device in Canada that was produced by SIEMENS HEALTHCARE LIMITED.

What is this?

A correction or removal action taken by a manufacturer to address a problem with a medical device. Recalls occur when a medical device is defective, when it could be a risk to health, or when it is both defective and a risk to health.

Learn more about the data here
  • Type of Event
    Recall
  • Event ID
    25438
  • Event Risk Class
    III
  • Event Initiated Date
    2017-09-26
  • Event Country
  • Event Source
    HC
  • Notes / Alerts
    Canadian data is current through March 2018. All of the data comes from Health Canada, except for the categories Manufacturer Parent Company and Product Classification.
    The Parent Company and the Product Classification were added by ICIJ.
    The parent company information is based on 2017 public records. The device classification information comes from FDA’s Product Classification by Review Panel, based on matches of recall data from the U.S. and Canada.
  • Extra notes in the data
  • Reason
    The wireless foot switch can fail due to impacts by external factors such as electrostatic discharge that exceeds a certain intensity. troubleshooting actions have been implemented (resetting the affected assembly) but have no effect in this case.

Device

  • Model / Serial
    Model Catalog: 4787797 (Lot serial: N/A); Model Catalog: 4787805 (Lot serial: N/A); Model Catalog: 4787813 (Lot serial: N/A)
  • Product Description
    WIRELESS XRAY FSW MONO;WIRELESS XRAY FSW BIPLANE;WIRELESS XRAY FSW MONO-MP
  • Manufacturer

Manufacturer

  • Manufacturer Address
    OAKVILLE
  • Manufacturer Parent Company (2017)
  • Source
    HC