Recall of VNS THERAPY DEMIPULSE SYSTEM - PROGRAMMING SOFTWARE

According to Health Canada (via FOI), this recall involved a device in Canada that was produced by XYCORP MEDICAL INC..

What is this?

A correction or removal action taken by a manufacturer to address a problem with a medical device. Recalls occur when a medical device is defective, when it could be a risk to health, or when it is both defective and a risk to health.

Learn more about the data here
  • Type of Event
    Recall
  • Event ID
    153153
  • Event Risk Class
    II
  • Event Initiated Date
    2009-08-11
  • Event Country
  • Event Source
    HC
  • Notes / Alerts
    Canadian data is current through March 2018. All of the data comes from Health Canada, except for the categories Manufacturer Parent Company and Product Classification.
    The Parent Company and the Product Classification were added by ICIJ.
    The parent company information is based on 2017 public records. The device classification information comes from FDA’s Product Classification by Review Panel, based on matches of recall data from the U.S. and Canada.
  • Extra notes in the data
  • Reason
    In rare instances a system diagnostic test using model 250 programming software (v 7.1 and earlier) may report "lead impedance: ok" when in fact normal wear and trauma has resulted in a short-circuit condition that prevents the lead from delivering therapy to the vagus nerve.

Device

  • Model / Serial
    Model Catalog: 250 (VERSIONS 6.1 AND LESS) (Lot serial: ALL VERSIONS AND LOT NUMBERS); Model Catalog: 250 V7.1.4 (Lot serial: ALL VERSIONS AND LOT NUMBERS); Model Catalog: 250 (Lot serial: ALL VERSIONS AND LOT NUMBERS)
  • Product Description
    VNS DEMIPULSE SYS PROGRAMMING SOFTWARE
  • Manufacturer

Manufacturer

  • Manufacturer Address
    MISSISSAUGA
  • Source