Recall of VITROS FS DILUENT PACK 3

According to Health Canada (via FOI), this recall involved a device in Canada that was produced by ORTHO-CLINICAL DIAGNOSTICS.

What is this?

A correction or removal action taken by a manufacturer to address a problem with a medical device. Recalls occur when a medical device is defective, when it could be a risk to health, or when it is both defective and a risk to health.

Learn more about the data here
  • Type of Event
    Recall
  • Event ID
    86942
  • Event Risk Class
    II
  • Event Initiated Date
    2014-03-31
  • Event Country
  • Event Source
    HC
  • Notes / Alerts
    Canadian data is current through March 2018. All of the data comes from Health Canada, except for the categories Manufacturer Parent Company and Product Classification.
    The Parent Company and the Product Classification were added by ICIJ.
    The parent company information is based on 2017 public records. The device classification information comes from FDA’s Product Classification by Review Panel, based on matches of recall data from the U.S. and Canada.
  • Extra notes in the data
  • Reason
    Ortho clinical diagnostics (ocd) internal testing has confirmed the potential for lower than expected vitros c-reactive protein (crp) results from samples diluted with vitros specialty diluent or vitros fs diluent pack 3. when using the affected lots ocd found that results using vitros crp slides may be negatively biased by approximately 51% using a 3x dilution factor.

Device

  • Model / Serial
    Model Catalog: 6801754 (Lot serial: 01-3266); Model Catalog: 855 9825 (Lot serial: F3168)
  • Product Description
    VITROS CHEMISTRY FS DILUENT PACK 3
  • Manufacturer

Manufacturer