Recall of VITROS ECI IMMUNODIAGNOSTIC SYSTEM - CA125II REAGENT PACK - CALIBRATOR

According to Health Canada (via FOI), this recall involved a device in Canada that was produced by ORTHO-CLINICAL DIAGNOSTICS.

What is this?

A correction or removal action taken by a manufacturer to address a problem with a medical device. Recalls occur when a medical device is defective, when it could be a risk to health, or when it is both defective and a risk to health.

Learn more about the data here
  • Type of Event
    Recall
  • Event ID
    119629
  • Event Risk Class
    II
  • Event Initiated Date
    2010-04-21
  • Event Country
  • Event Source
    HC
  • Notes / Alerts
    Canadian data is current through March 2018. All of the data comes from Health Canada, except for the categories Manufacturer Parent Company and Product Classification.
    The Parent Company and the Product Classification were added by ICIJ.
    The parent company information is based on 2017 public records. The device classification information comes from FDA’s Product Classification by Review Panel, based on matches of recall data from the U.S. and Canada.
  • Extra notes in the data
  • Reason
    There is a potential for negative biases when testing patient samples. an observed negative bias in patient samples could mask moderate increases in ca 125 which could potentially delay the use of additional diagnostic and therapeutic measures used to verify and treat disease relapses or progression. qc samples do not exhibit the bais observed with patient samples.