Recall of VITROS CHEMISTRY PRODUCTS SYSTEM - 250/250AT CHEMISTRY ANALYZER - CLASS 3

According to Health Canada (via FOI), this recall involved a device in Canada that was produced by ORTHO-CLINICAL DIAGNOSTICS.

What is this?

A correction or removal action taken by a manufacturer to address a problem with a medical device. Recalls occur when a medical device is defective, when it could be a risk to health, or when it is both defective and a risk to health.

Learn more about the data here
  • Type of Event
    Recall
  • Event ID
    43377
  • Event Risk Class
    III
  • Event Initiated Date
    2015-10-01
  • Event Country
  • Event Source
    HC
  • Notes / Alerts
    Canadian data is current through March 2018. All of the data comes from Health Canada, except for the categories Manufacturer Parent Company and Product Classification.
    The Parent Company and the Product Classification were added by ICIJ.
    The parent company information is based on 2017 public records. The device classification information comes from FDA’s Product Classification by Review Panel, based on matches of recall data from the U.S. and Canada.
  • Extra notes in the data
  • Reason
    Ortho clinical diagnostics inc. (ocd) initiated this recall due to the potential for an increase in u90-382 or 6lu condition codes (wash errors) when using vitros calibrator kit 9 lot 0954. when a u90-382 or 6lu condition code occurs the vitros system suppresses the result and no result is reported. the increase in condition codes primarily occurs when using vitros dgxn and phyt slides. all numerical results obtained are valid providing that no wash errors occurred and quality control results were within acceptable limits.

Device

  • Model / Serial
    Model Catalog: 8568040 (Lot serial: 0954)
  • Product Description
    VITROS Chemistry Products Calibrator Kit 9
  • Manufacturer

Manufacturer