Recall of VIAL-MATE ADAPTOR

According to Health Canada (via FOI), this recall involved a device in Canada that was produced by BAXTER CORPORATION.

What is this?

A correction or removal action taken by a manufacturer to address a problem with a medical device. Recalls occur when a medical device is defective, when it could be a risk to health, or when it is both defective and a risk to health.

Learn more about the data here
  • Type of Event
    Recall
  • Event ID
    70182
  • Event Risk Class
    II
  • Event Initiated Date
    2010-12-06
  • Event Country
  • Event Source
    HC
  • Notes / Alerts
    Canadian data is current through March 2018. All of the data comes from Health Canada, except for the categories Manufacturer Parent Company and Product Classification.
    The Parent Company and the Product Classification were added by ICIJ.
    The parent company information is based on 2017 public records. The device classification information comes from FDA’s Product Classification by Review Panel, based on matches of recall data from the U.S. and Canada.
  • Extra notes in the data
  • Reason
    Baxter has become aware that the vial-mate vial reconstitution device is not fully effective when used with i.V. bags that do not contain a minimum amount of air. the viaflex bags currently marketed in canada are not injected with air during production and may not contain this minimum amount of air. therefore incomplete reconstitution may result when the vial-mate device is used with the viaflex bags.

Device

  • Model / Serial
    Model Catalog: 2B8071 (Lot serial: >10 NUMBERS CONTACT MFR)
  • Product Description
    VIAL-MATE ADAPTOR
  • Manufacturer

Manufacturer