Recall of VERIS 8600 VITAL SIGNS MONITOR - BASE SYSTEM

According to Health Canada (via FOI), this recall involved a device in Canada that was produced by CSP MEDICAL.

What is this?

A correction or removal action taken by a manufacturer to address a problem with a medical device. Recalls occur when a medical device is defective, when it could be a risk to health, or when it is both defective and a risk to health.

Learn more about the data here
  • Type of Event
    Recall
  • Event ID
    24452
  • Event Risk Class
    II
  • Event Initiated Date
    2013-05-05
  • Event Country
  • Event Source
    HC
  • Notes / Alerts
    Canadian data is current through March 2018. All of the data comes from Health Canada, except for the categories Manufacturer Parent Company and Product Classification.
    The Parent Company and the Product Classification were added by ICIJ.
    The parent company information is based on 2017 public records. The device classification information comes from FDA’s Product Classification by Review Panel, based on matches of recall data from the U.S. and Canada.
  • Extra notes in the data
  • Reason
    Two complaints were received reporting the dc power cable (p/n 3010556) jacket attached to the veris unit melted. the root cause investigation determined a design reliability issue at one of the cable terminations was found to be a contributing factor since it allows the wires at the end of the cables to be bent and twisted during use which can lead to electrical shorting.

Device

  • Model / Serial
    Model Catalog: 3011992 (Lot serial: N/A); Model Catalog: 3011993 (Lot serial: N/A); Model Catalog: 3014268 (Lot serial: N/A); Model Catalog: 3011994 (Lot serial: N/A); Model Catalog: 3011996 (Lot serial: N/A)
  • Product Description
    VERIS 8600 VITAL SIGNS MONITOR - BASE SYSTEM
  • Manufacturer

Manufacturer

  • Manufacturer Address
    LONDON
  • Manufacturer Parent Company (2017)
  • Source
    HC