Recall of VARIS TREATMENT

According to Health Canada (via FOI), this recall involved a device in Canada that was produced by VARIAN MEDICAL SYSTEMS INC..

What is this?

A correction or removal action taken by a manufacturer to address a problem with a medical device. Recalls occur when a medical device is defective, when it could be a risk to health, or when it is both defective and a risk to health.

Learn more about the data here
  • Type of Event
    Recall
  • Event ID
    65263
  • Event Risk Class
    II
  • Event Initiated Date
    2008-11-22
  • Event Country
  • Event Source
    HC
  • Notes / Alerts
    Canadian data is current through March 2018. All of the data comes from Health Canada, except for the categories Manufacturer Parent Company and Product Classification.
    The Parent Company and the Product Classification were added by ICIJ.
    The parent company information is based on 2017 public records. The device classification information comes from FDA’s Product Classification by Review Panel, based on matches of recall data from the U.S. and Canada.
  • Extra notes in the data
  • Reason
    A malfunction in varis software if undetected could result in an under dose of as much as 100%. commas used instead of decimals for numerical separators with this version 6.6.5042 are not understood properly.

Device

  • Model / Serial
    Model Catalog: PRODUCT CODE: H46 (Lot serial: H462021); Model Catalog: VERSION 6.6.5042 6.6.504 (Lot serial: H462022); Model Catalog: PRODUCT CODE: H46 (Lot serial: H462022); Model Catalog: VERSION 6.6.5042 6.6.504 (Lot serial: H462023); Model Catalog: PRODUCT CODE: H46 (Lot serial: H462023); Model Catalog: VERSION 6.6.5042 6.6.504 (Lot serial: H462024); Model Catalog: PRODUCT CODE: H46 (Lot serial: H462024); Model Catalog: PRODUCT CODE: H46 (Lot serial: H462017); Model Catalog: VERSION 6.6.5042 6.6.504 (Lot serial: H462017); Model Catalog: PRODUCT CODE: H46 (Lot serial: H462016); Model Catalog: VERSION 6.6.5042 6.6.504 (Lot serial: H462016); Model Catalog: VERSION 6.6.5042 6.6.504 (Lot serial: H462021)
  • Product Description
    Varian Medical Varis Treatment System
  • Manufacturer

Manufacturer