Recall of UNIVERSAL SLING BAR

According to Health Canada (via FOI), this recall involved a device in Canada that was produced by HILL-ROM CANADA LTD..

What is this?

A correction or removal action taken by a manufacturer to address a problem with a medical device. Recalls occur when a medical device is defective, when it could be a risk to health, or when it is both defective and a risk to health.

Learn more about the data here
  • Type of Event
    Recall
  • Event ID
    101418
  • Event Risk Class
    II
  • Event Initiated Date
    2016-01-05
  • Event Country
  • Event Source
    HC
  • Notes / Alerts
    Canadian data is current through March 2018. All of the data comes from Health Canada, except for the categories Manufacturer Parent Company and Product Classification.
    The Parent Company and the Product Classification were added by ICIJ.
    The parent company information is based on 2017 public records. The device classification information comes from FDA’s Product Classification by Review Panel, based on matches of recall data from the U.S. and Canada.
  • Extra notes in the data
  • Reason
    It has been determined that the sling bar bolt can be subjected to bending throughout its lifetime when used outside the expected user profile those misuses include lifting repeatedly with only one hook (a single side of the sling bar) lifting at angles using one hook of the lift to pull patients in bed or even using the sling bar to lift heavy equipment for servicing activities.

Device

  • Model / Serial
    Model Catalog: 3156076 (Lot serial: 1200101 to 1370151); Model Catalog: 3156095 (Lot serial: 1200101 to 1370151); Model Catalog: 3156094 (Lot serial: 1200101 to 1370151); Model Catalog: 3156075 (Lot serial: 1200101 to 1370151); Model Catalog: 3156074 (Lot serial: 1200101 to 1370151); Model Catalog: 3156086 (Lot serial: 1200101 to 1370151); Model Catalog: 3156085 (Lot serial: 1200101 to 1370151); Model Catalog: 3156084 (Lot serial: 1200101 to 1370151)
  • Product Description
    UNIVERSAL SLING BAR
  • Manufacturer