Recall of UNIVERSAL SLING BAR

According to Health Canada (via FOI), this recall involved a device in Canada that was produced by LIKO AB.

What is this?

A correction or removal action taken by a manufacturer to address a problem with a medical device. Recalls occur when a medical device is defective, when it could be a risk to health, or when it is both defective and a risk to health.

Learn more about the data here
  • Type of Event
    Recall
  • Event ID
    54787
  • Event Risk Class
    II
  • Event Initiated Date
    2008-07-11
  • Event Country
  • Event Source
    HC
  • Notes / Alerts
    Canadian data is current through March 2018. All of the data comes from Health Canada, except for the categories Manufacturer Parent Company and Product Classification.
    The Parent Company and the Product Classification were added by ICIJ.
    The parent company information is based on 2017 public records. The device classification information comes from FDA’s Product Classification by Review Panel, based on matches of recall data from the U.S. and Canada.
  • Extra notes in the data
  • Reason
    When the universal slingbar is mounted on viking or likolight mobile lifts a component securing the universal slingbar may experience unanticipated fatigue which in extreme cases can result in the sling bar.

Device

  • Model / Serial
    Model Catalog: 3156086 (Lot serial: 121204 to 121499); Model Catalog: 3156075 (Lot serial: 1200101 to 1245680); Model Catalog: 3156076 (Lot serial: 1200101 to 1245680); Model Catalog: 3156084 (Lot serial: 1200101 to 1245680); Model Catalog: 3156085 (Lot serial: 1200101 to 1245680); Model Catalog: 3156086 (Lot serial: 12222407 to 12225241); Model Catalog: 3156085 (Lot serial: 12222407 to 12225241); Model Catalog: 3156084 (Lot serial: 12222407 to 12225241); Model Catalog: 3156076 (Lot serial: 12222407 to 12225241); Model Catalog: 3156075 (Lot serial: 12222407 to 12225241); Model Catalog: 3156074 (Lot serial: 12222407 to 12225241); Model Catalog: 3156074 (Lot serial: 1200101 to 1245680); Model Catalog: 3156085 (Lot serial: 121204 to 121499); Model Catalog: 3156084 (Lot serial: 121204 to 121499); Model Catalog: 3156076 (Lot serial: 121204 to 121499); Model Catalog: 3156075 (Lot serial: 121204 to 121499); Model Catalog: 3156074 (Lot serial: 121204 to 121499); Model Catalog: 3156086 (Lot serial: 1200101 to 1245680
  • Product Description
    Universal SlingBar 350/450/600
  • Manufacturer

Manufacturer