Recall of UNICEL DXH SLIDEMAKER STAINER

According to Health Canada (via FOI), this recall involved a device in Canada that was produced by BECKMAN COULTER CANADA L.P..

What is this?

A correction or removal action taken by a manufacturer to address a problem with a medical device. Recalls occur when a medical device is defective, when it could be a risk to health, or when it is both defective and a risk to health.

Learn more about the data here
  • Type of Event
    Recall
  • Event ID
    85742
  • Event Risk Class
    III
  • Event Initiated Date
    2014-02-12
  • Event Country
  • Event Source
    HC
  • Notes / Alerts
    Canadian data is current through March 2018. All of the data comes from Health Canada, except for the categories Manufacturer Parent Company and Product Classification.
    The Parent Company and the Product Classification were added by ICIJ.
    The parent company information is based on 2017 public records. The device classification information comes from FDA’s Product Classification by Review Panel, based on matches of recall data from the U.S. and Canada.
  • Extra notes in the data
  • Reason
    If the "ordering physician" option is selected in the system manager label configuration screen the unicel dxh slidemaker stainer coulter cellular analysis system will print the name of the primary physician instead of the ordering physician on the slide label. the ordering physician's name correctly transmits to the laboratory information system. the primary identifier the date/time and other demographic information will appear as selected on the slide label. slide making and staining quality are not affected.

Device

  • Model / Serial
    Model Catalog: 775222 (Lot serial: >10 NUMBERS CONTACT MFR)
  • Product Description
    UNICEL DXH SLIDEMAKER STAINER SYSTEM
  • Manufacturer

Manufacturer