  • Reason
    Memory allocation error (0x13890001): accumulation of message manager tasks uses up available dxc console memory causing the console to shutdown.

Device

  • Model / Serial
    Model Catalog: A59102 (Lot serial: 4143); Model Catalog: A11812 (Lot serial: 2113); Model Catalog: A11812 (Lot serial: 2337)
  • Product Description
    UNICEL DXC 800I
  • Manufacturer

Manufacturer