Recall of UNICEL DXC 600I SYNCHRON ACCESS CLINICAL ANALYZER

According to Health Canada (via FOI), this recall involved a device in Canada that was produced by BECKMAN COULTER CANADA L.P..

What is this?

A correction or removal action taken by a manufacturer to address a problem with a medical device. Recalls occur when a medical device is defective, when it could be a risk to health, or when it is both defective and a risk to health.

Learn more about the data here
  • Type of Event
    Recall
  • Event ID
    80275
  • Event Risk Class
    III
  • Event Initiated Date
    2012-05-03
  • Event Country
  • Event Source
    HC
  • Notes / Alerts
    Canadian data is current through March 2018. All of the data comes from Health Canada, except for the categories Manufacturer Parent Company and Product Classification.
    The Parent Company and the Product Classification were added by ICIJ.
    The parent company information is based on 2017 public records. The device classification information comes from FDA’s Product Classification by Review Panel, based on matches of recall data from the U.S. and Canada.
  • Extra notes in the data
  • Reason
    Access testosterone reagent kits have included document a82247a ("kit stuffer") since december 2009. the kit stuffer incorrectly states that unicel dxc600i instruments are not implicated in a product corrective action regarding reagent contamination. dxc600i customers were informed of the reagent contamination issue in october 2007 in a recall letter (pca-c-e-1051) but the specific reference to the dxc600i instrument was inadvertently omitted from the testosterone kit stuffer.

Device

  • Model / Serial
    Model Catalog: 4767 (Lot serial: =<800521 CONTACT MFG)
  • Product Description
    ACCESS TESTOSTERONE - UNICEL DXC 600I SYNCHRON ACCESS CLINICAL SYSTEMS