Recall of ULTRAVIEW SL COMPACT MONITOR

According to Health Canada (via FOI), this recall involved a device in Canada that was produced by SPACELABS HEALTHCARE (CANADA) INC..

What is this?

A correction or removal action taken by a manufacturer to address a problem with a medical device. Recalls occur when a medical device is defective, when it could be a risk to health, or when it is both defective and a risk to health.

Learn more about the data here
  • Type of Event
    Recall
  • Event ID
    78301
  • Event Risk Class
    II
  • Event Initiated Date
    2009-11-13
  • Event Country
  • Event Source
    HC
  • Notes / Alerts
    Canadian data is current through March 2018. All of the data comes from Health Canada, except for the categories Manufacturer Parent Company and Product Classification.
    The Parent Company and the Product Classification were added by ICIJ.
    The parent company information is based on 2017 public records. The device classification information comes from FDA’s Product Classification by Review Panel, based on matches of recall data from the U.S. and Canada.
  • Extra notes in the data
  • Reason
    Affected spacelabs' portable patient monitors may stop working when on ac power or shut down and do not power on again. the failure mode is caused by the capacitor identified as c329 malfunctioning and initiating a monitor shutdown.

Device

  • Model / Serial
    Model Catalog: 91369 (Lot serial: >10 NUMBERS CONTACT MFR); Model Catalog: 91367 (Lot serial: 1367-000449); Model Catalog: 91367 (Lot serial: 1367-000447); Model Catalog: 91367 (Lot serial: 1367-000488); Model Catalog: 91367 (Lot serial: 1367-000484); Model Catalog: 91367 (Lot serial: 1367-000448); Model Catalog: 91370 (Lot serial: >10 NUMBERS CONTACT MFR)
  • Product Description
    ULTRAVIEW SL COMPACT MONITOR
  • Manufacturer

Manufacturer

  • Manufacturer Address
    MISSISSAUGA
  • Manufacturer Parent Company (2017)
  • Source
    HC