Recall of ULTRAVIEW SL BEDSIDE/CENTRAL MONITOR

According to Health Canada (via FOI), this recall involved a device in Canada that was produced by SPACELABS HEALTHCARE (CANADA) INC..

What is this?

A correction or removal action taken by a manufacturer to address a problem with a medical device. Recalls occur when a medical device is defective, when it could be a risk to health, or when it is both defective and a risk to health.

Learn more about the data here
  • Type of Event
    Recall
  • Event ID
    51327
  • Event Risk Class
    II
  • Event Initiated Date
    2011-04-09
  • Event Country
  • Event Source
    HC
  • Notes / Alerts
    Canadian data is current through March 2018. All of the data comes from Health Canada, except for the categories Manufacturer Parent Company and Product Classification.
    The Parent Company and the Product Classification were added by ICIJ.
    The parent company information is based on 2017 public records. The device classification information comes from FDA’s Product Classification by Review Panel, based on matches of recall data from the U.S. and Canada.
  • Extra notes in the data
  • Reason
    The manufacturer has determined through internal testing that the circuit that supplies backup power to the modules is not working. if there is a power interruption the 3-minute backup may not be operational and you may lose any changes you have made to the alarm settings. when used with model 91387 the hazard presents a risk for all modules except modules 90470 90478 90496 & 91496 all of which have their own internal backup and will not lose changes to the alarm settings for power interruptions less than 3 minutes.

Device

  • Model / Serial
    Model Catalog: 91387 (Lot serial: >100 NUMBERS CONTACT MFR)
  • Product Description
    SPACELABS MEDICAL PATIENT MONITOR91387