Recall of TOTALCARE BED SYSTEM - BASE FRAME

According to Health Canada (via FOI), this recall involved a device in Canada that was produced by HILL-ROM CANADA LTD..

What is this?

A correction or removal action taken by a manufacturer to address a problem with a medical device. Recalls occur when a medical device is defective, when it could be a risk to health, or when it is both defective and a risk to health.

Learn more about the data here
  • Type of Event
    Recall
  • Event ID
    67267
  • Event Risk Class
    II
  • Event Initiated Date
    2017-02-17
  • Event Country
  • Event Source
    HC
  • Notes / Alerts
    Canadian data is current through March 2018. All of the data comes from Health Canada, except for the categories Manufacturer Parent Company and Product Classification.
    The Parent Company and the Product Classification were added by ICIJ.
    The parent company information is based on 2017 public records. The device classification information comes from FDA’s Product Classification by Review Panel, based on matches of recall data from the U.S. and Canada.
  • Extra notes in the data
  • Reason
    Hill rom became aware that the bed exit alarm on totalcare beds equipped with the 3 level patient position monitoring may not function properly if the alarm has been armed for a period longer thank seventy-two (72) hours without the alarm being triggered by patient movement or the patient exiting the bed within that that time period.

Device

  • Model / Serial
    Model Catalog: P1900 (Lot serial: >10 see mfr)
  • Product Description
    TOTALCARE BED SYSTEM - BASE FRAME
  • Manufacturer

Manufacturer