Recall of TOMOTHERAPY TREATMENT SYSTEM - MAIN UNIT

According to Health Canada (via FOI), this recall involved a device in Canada that was produced by ACCURAY INCORPORATED.

What is this?

A correction or removal action taken by a manufacturer to address a problem with a medical device. Recalls occur when a medical device is defective, when it could be a risk to health, or when it is both defective and a risk to health.

Learn more about the data here
  • Type of Event
    Recall
  • Event ID
    48077
  • Event Risk Class
    II
  • Event Initiated Date
    2011-01-11
  • Event Country
  • Event Source
    HC
  • Notes / Alerts
    Canadian data is current through March 2018. All of the data comes from Health Canada, except for the categories Manufacturer Parent Company and Product Classification.
    The Parent Company and the Product Classification were added by ICIJ.
    The parent company information is based on 2017 public records. The device classification information comes from FDA’s Product Classification by Review Panel, based on matches of recall data from the U.S. and Canada.
  • Extra notes in the data
  • Reason
    With some specific plan types the treatment planning system can over predict dose to targets resulting in potential under dose. the plan elements necessary to encounter this issue are sharp dose gradients with few simple and small axial cross section targets located off central axis.

Device

  • Model / Serial
    Model Catalog: H-0000-0003 (Lot serial: >10 NUMBERS CONTACT MFG)
  • Product Description
    TOMOTHERAPY TREATMENT SYSTEM - MAIN UNIT
  • Manufacturer

Manufacturer

  • Manufacturer Address
    MADISON
  • Manufacturer Parent Company (2017)
  • Source
    HC