Recall of T2100 TREADMILL

According to Health Canada (via FOI), this recall involved a device in Canada that was produced by GENERAL ELECTRIC CANADA (OPERATING AS GE HEALTHCARE).

What is this?

A correction or removal action taken by a manufacturer to address a problem with a medical device. Recalls occur when a medical device is defective, when it could be a risk to health, or when it is both defective and a risk to health.

Learn more about the data here
  • Type of Event
    Recall
  • Event ID
    69662
  • Event Risk Class
    II
  • Event Initiated Date
    2015-02-10
  • Event Country
  • Event Source
    HC
  • Notes / Alerts
    Canadian data is current through March 2018. All of the data comes from Health Canada, except for the categories Manufacturer Parent Company and Product Classification.
    The Parent Company and the Product Classification were added by ICIJ.
    The parent company information is based on 2017 public records. The device classification information comes from FDA’s Product Classification by Review Panel, based on matches of recall data from the U.S. and Canada.
  • Extra notes in the data
  • Reason
    Manufacturer internal quality inspection has found that the power cord connecting directly to the t2100 treadmill may not have been assembled according to specifications. if the power cord was improperly assembled and a separate secondary electrical fault condition exists (e.G. a frayed extension or power cord touching the treadmill chassis) this could possibly result in an electrical shock to the patient or operator. there have been no reported incidences of the treadmill power cord leading to an electrical shock of a patient or operator at this time.

Device

Manufacturer