Recall of SYSTEM DI-60 MAIN UNIT

According to Health Canada (via FOI), this recall involved a device in Canada that was produced by SYSMEX CANADA INC..

What is this?

A correction or removal action taken by a manufacturer to address a problem with a medical device. Recalls occur when a medical device is defective, when it could be a risk to health, or when it is both defective and a risk to health.

Learn more about the data here
  • Type of Event
    Recall
  • Event ID
    48337
  • Event Risk Class
    III
  • Event Initiated Date
    2016-09-14
  • Event Country
  • Event Source
    HC
  • Notes / Alerts
    Canadian data is current through March 2018. All of the data comes from Health Canada, except for the categories Manufacturer Parent Company and Product Classification.
    The Parent Company and the Product Classification were added by ICIJ.
    The parent company information is based on 2017 public records. The device classification information comes from FDA’s Product Classification by Review Panel, based on matches of recall data from the U.S. and Canada.
  • Extra notes in the data
  • Reason
    There are sudden failures of the di-60 home position sensors that cause multiple error messages and prevents operation of the system. the home position sensors enable the analyzer to recognize proper positioning of the slides the camera and optics to scan and locate cells on a stained blood smear. failure of any one of the sensors disables the analyzer.

Device

  • Model / Serial
    Model Catalog: CC286297 (Lot serial: SERIAL RANGE 60020 - 60155)
  • Product Description
    AUTOMATED DIGITAL CELL MORPHOLOGY ANALYZER DI-60
  • Manufacturer

Manufacturer

  • Manufacturer Address
    MISSISSAUGA
  • Manufacturer Parent Company (2017)
  • Source
    HC