Recall of SYSMEX POCH-100I ANALYZER

According to Health Canada (via FOI), this recall involved a device in Canada that was produced by SYSMEX CANADA INC..

What is this?

A correction or removal action taken by a manufacturer to address a problem with a medical device. Recalls occur when a medical device is defective, when it could be a risk to health, or when it is both defective and a risk to health.

Learn more about the data here
  • Type of Event
    Recall
  • Event ID
    151194
  • Event Risk Class
    II
  • Event Initiated Date
    2013-07-12
  • Event Country
  • Event Source
    HC
  • Notes / Alerts
    Canadian data is current through March 2018. All of the data comes from Health Canada, except for the categories Manufacturer Parent Company and Product Classification.
    The Parent Company and the Product Classification were added by ICIJ.
    The parent company information is based on 2017 public records. The device classification information comes from FDA’s Product Classification by Review Panel, based on matches of recall data from the U.S. and Canada.
  • Extra notes in the data
  • Reason
    In rare instances the data printed on the thermal printer may omit a digit or decimal point in the results or sample identification number. this is isolated to the thermal printout and does not affect the values displayed on the lcd screen graphic printer or output to the host computer.

Device

  • Model / Serial
    Model Catalog: 023-1861-8 (Lot serial: >10 NUMBERS CONTACT MFG)
  • Product Description
    SYSMEX POCH-100I AUTOMATED HEMATOLOGY ANALYZER
  • Manufacturer

Manufacturer

  • Manufacturer Address
    MISSISSAUGA
  • Manufacturer Parent Company (2017)
  • Source
    HC