Recall of SYNCHRON LXI 725

According to Health Canada (via FOI), this recall involved a device in Canada that was produced by BECKMAN COULTER CANADA L.P..

What is this?

A correction or removal action taken by a manufacturer to address a problem with a medical device. Recalls occur when a medical device is defective, when it could be a risk to health, or when it is both defective and a risk to health.

Learn more about the data here
  • Type of Event
    Recall
  • Event ID
    25443
  • Event Risk Class
    III
  • Event Initiated Date
    2012-03-13
  • Event Country
  • Event Source
    HC
  • Notes / Alerts
    Canadian data is current through March 2018. All of the data comes from Health Canada, except for the categories Manufacturer Parent Company and Product Classification.
    The Parent Company and the Product Classification were added by ICIJ.
    The parent company information is based on 2017 public records. The device classification information comes from FDA’s Product Classification by Review Panel, based on matches of recall data from the U.S. and Canada.
  • Extra notes in the data
  • Reason
    The incubator belt (pn 61-007256-4) used on all access 2 immunoassay systems made from 2008 to feb 2012 does not conform to specifications. one tooth per belt is wider than the specification and can shear off from the incubator belt which may cause a jam or motion error. this may lead to a delay in the reporting of results.

Device

  • Model / Serial
    Model Catalog: 476501 (Lot serial: ALL); Model Catalog: 4767 (Lot serial: ALL); Model Catalog: 81600N (Lot serial: ALL)
  • Product Description
    SYNCHRON LXI 725 CLINICAL SYSTEM INCUBATOR BELT
  • Manufacturer

Manufacturer