Recall of SYNCHROMED II - PROGRAMMABLE PUMP

According to Health Canada (via FOI), this recall involved a device in Canada that was produced by MEDTRONIC OF CANADA LTD..

What is this?

A correction or removal action taken by a manufacturer to address a problem with a medical device. Recalls occur when a medical device is defective, when it could be a risk to health, or when it is both defective and a risk to health.

Learn more about the data here
  • Type of Event
    Recall
  • Event ID
    43454
  • Event Risk Class
    II
  • Event Initiated Date
    2014-03-05
  • Event Country
  • Event Source
    HC
  • Notes / Alerts
    Canadian data is current through March 2018. All of the data comes from Health Canada, except for the categories Manufacturer Parent Company and Product Classification.
    The Parent Company and the Product Classification were added by ICIJ.
    The parent company information is based on 2017 public records. The device classification information comes from FDA’s Product Classification by Review Panel, based on matches of recall data from the U.S. and Canada.
  • Extra notes in the data
  • Reason
    Provide important new information regarding overinfusion associated with the synchromed ii implantable pump.Overinfusion is defined as an infusion rate exceeding the programmed infusion rate by more than 14.5% as described in the labeling. when overinfusion occurs it will result in a volume discrepancy at pump refill where the volume withdrawn from the pump is less than the volume expected. overinfusion can result in a life-threatening overdose and can also result in drug withdrawal due to premature emptying of the pump. due to the low reported rate of occurrence of this issue and the inability to predict which pumps may be at risk medtronic is not recommending prophylactic replacement of pumps.

Device

  • Model / Serial
    Model Catalog: 8637 (Lot serial: All Lots)
  • Product Description
    SYNCHROMED II - PROGRAMMABLE PUMP
  • Manufacturer

Manufacturer

  • Manufacturer Address
    BRAMPTON
  • Manufacturer Parent Company (2017)
  • Manufacturer comment
    “If our surveillance systems identify a potential performance issue, our personnel promptly evaluate the problem, including, when appropriate, conducting root cause investigations and internal testing to assess whether the product continues to meet specifications and defined performance criteria,” Medtronic told ICIJ in a statement. “In some cases, based on this evaluation, Medtronic may determine that a recall is necessary.” The company said that it communicates with healthcare providers and/or patients and provide recommendations to address such issues. Medtronic noted that these communications can include letters, emails, calls, press releases, physician notifications and social media postings, as well as informing the FDA and other regulators of the actions.