Recall of SYMBIQ

According to Health Canada (via FOI), this recall involved a device in Canada that was produced by HOSPIRA HEALTHCARE CORPORATION.

What is this?

A correction or removal action taken by a manufacturer to address a problem with a medical device. Recalls occur when a medical device is defective, when it could be a risk to health, or when it is both defective and a risk to health.

Learn more about the data here
  • Type of Event
    Recall
  • Event ID
    65093
  • Event Risk Class
    II
  • Event Initiated Date
    2011-10-18
  • Event Country
  • Event Source
    HC
  • Notes / Alerts
    Canadian data is current through March 2018. All of the data comes from Health Canada, except for the categories Manufacturer Parent Company and Product Classification.
    The Parent Company and the Product Classification were added by ICIJ.
    The parent company information is based on 2017 public records. The device classification information comes from FDA’s Product Classification by Review Panel, based on matches of recall data from the U.S. and Canada.
    ICU Medical, Inc. acquired from Pfizer in early 2017, Hospira Infusion Systems, the portion of Hospira dedicated to develop infusion pumps.
  • Extra notes in the data
  • Reason
    In instances where the pump alarms for accumulated air in line and detects froth in quick sucession the infusion will stop as desired but a malfunction might occur where the pump will stop operating.

Device

  • Model / Serial
    Model Catalog: 16027 (Lot serial: ); Model Catalog: 16026 (Lot serial: )
  • Product Description
    Symbiq Infuser
  • Manufacturer

Manufacturer

  • Manufacturer Address
    SAINT-LAURENT
  • Manufacturer Parent Company (2017)
  • Source
    HC