Recall of STRATUS CS ACUTE CARE SYSTEM - D-DIMER (DDMR) ASSAY

According to Health Canada (via FOI), this recall involved a device in Canada that was produced by SIEMENS HEALTHCARE LIMITED.

What is this?

A correction or removal action taken by a manufacturer to address a problem with a medical device. Recalls occur when a medical device is defective, when it could be a risk to health, or when it is both defective and a risk to health.

Learn more about the data here
  • Type of Event
    Recall
  • Event ID
    132294
  • Event Risk Class
    III
  • Event Initiated Date
    2012-06-11
  • Event Country
  • Event Source
    HC
  • Notes / Alerts
    Canadian data is current through March 2018. All of the data comes from Health Canada, except for the categories Manufacturer Parent Company and Product Classification.
    The Parent Company and the Product Classification were added by ICIJ.
    The parent company information is based on 2017 public records. The device classification information comes from FDA’s Product Classification by Review Panel, based on matches of recall data from the U.S. and Canada.
  • Extra notes in the data
  • Reason
    May experience calibration failure. the stratus cs instrument trigger the following error codes: slope and calcv error. calibrations that result in a failure prevents the d-dimer assay from being run on the stratus cs instrument until a successful calibration can be achieved.

Device

  • Model / Serial
    Model Catalog: CDDMR-C (Lot serial: 502058002); Model Catalog: CDDMR-C (Lot serial: 501318002)
  • Product Description
    STRATUS CS ACUTE CARE SYSTEM-D-DIMER (DDMR) ASSAY
  • Manufacturer

Manufacturer

  • Manufacturer Address
    OAKVILLE
  • Manufacturer Parent Company (2017)
  • Source
    HC