Recall of STOCKERT HEATER COOLER SYSTEM 3T - MAIN UNIT

According to Health Canada (via FOI), this recall involved a device in Canada that was produced by LIVANOVA CANADA CORP..

What is this?

A correction or removal action taken by a manufacturer to address a problem with a medical device. Recalls occur when a medical device is defective, when it could be a risk to health, or when it is both defective and a risk to health.

Learn more about the data here
  • Type of Event
    Recall
  • Event ID
    54626
  • Event Risk Class
    II
  • Event Initiated Date
    2017-08-15
  • Event Country
  • Event Source
    HC
  • Notes / Alerts
    Canadian data is current through March 2018. All of the data comes from Health Canada, except for the categories Manufacturer Parent Company and Product Classification.
    The Parent Company and the Product Classification were added by ICIJ.
    The parent company information is based on 2017 public records. The device classification information comes from FDA’s Product Classification by Review Panel, based on matches of recall data from the U.S. and Canada.
  • Extra notes in the data
  • Reason
    This is a follow-up to previous field safety notices issued in june 2015 and october 2016. livanova has developed a solution to mitigate the dispersion of aerosols produced by the 3t heater-cooler (3t). the solution is a retrofit upgrade of devices currently in use at facilities and at livanova to reduce and mitigate potential risk of transmission of nontuberculous mycobacteria (ntm) due to aerosolization.

Device

  • Model / Serial
    Model Catalog: 16-02-85 (Lot serial: >100 NUMBERS (CONTACT MFR)); Model Catalog: 16-02-81 (Lot serial: >100 NUMBERS (CONTACT MFR)); Model Catalog: 16-02-82 (Lot serial: >100 NUMBERS (CONTACT MFR))
  • Product Description
    STOCKERT HEATER-COOLER SYSTEM 3T
  • Manufacturer

Manufacturer

  • Manufacturer Address
    MARKHAM
  • Manufacturer Parent Company (2017)
  • Source
    HC