Recall of STERILMED REPROCESSED STRYKER EXTERNAL FIXATION SYSTEMS

According to Health Canada (via FOI), this recall involved a device in Canada that was produced by JOHNSON & JOHNSON MEDICAL PRODUCTS A DIV. OF JOHNSON & JOHNSON INC..

What is this?

A correction or removal action taken by a manufacturer to address a problem with a medical device. Recalls occur when a medical device is defective, when it could be a risk to health, or when it is both defective and a risk to health.

Learn more about the data here
  • Type of Event
    Recall
  • Event ID
    31533
  • Event Risk Class
    III
  • Event Initiated Date
    2017-10-26
  • Event Country
  • Event Source
    HC
  • Notes / Alerts
    Canadian data is current through March 2018. All of the data comes from Health Canada, except for the categories Manufacturer Parent Company and Product Classification.
    The Parent Company and the Product Classification were added by ICIJ.
    The parent company information is based on 2017 public records. The device classification information comes from FDA’s Product Classification by Review Panel, based on matches of recall data from the U.S. and Canada.
  • Extra notes in the data
  • Reason
    Sterilmed has initiated a medical device correction for all lots of listed sterilmed reprocessed depuy synthes stryker and smith & nephew external fixation systems. sterilmed has discovered the product label and ifu do not contain information regarding the safety and compatibility of the subject products in a magnetic resonance environment and such labeling is inconsistent with the oem labeling.

Device

  • Model / Serial
    Model Catalog: STR5048-5-150 (Lot serial: ALL LOTS); Model Catalog: STR5048-5-100 (Lot serial: ALL LOTS); Model Catalog: STR5029-8-840 (Lot serial: ALL LOTS); Model Catalog: STR5029-8-835 (Lot serial: ALL LOTS); Model Catalog: STR5029-8-830 (Lot serial: ALL LOTS); Model Catalog: STR5029-8-825 (Lot serial: ALL LOTS); Model Catalog: STR5029-8-820 (Lot serial: ALL LOTS); Model Catalog: STR5029-8-815 (Lot serial: ALL LOTS); Model Catalog: STR5029-8-810 (Lot serial: ALL LOTS); Model Catalog: STR5029-8-805 (Lot serial: ALL LOTS); Model Catalog: STR5029-8-640 (Lot serial: ALL LOTS); Model Catalog: STR5029-8-635 (Lot serial: ALL LOTS); Model Catalog: STR5029-8-630 (Lot serial: ALL LOTS); Model Catalog: STR5029-8-625 (Lot serial: ALL LOTS); Model Catalog: STR5029-8-620 (Lot serial: ALL LOTS); Model Catalog: STR5029-8-615 (Lot serial: ALL LOTS); Model Catalog: STR5029-8-610 (Lot serial: ALL LOTS); Model Catalog: STR5029-8-605 (Lot serial: ALL LOTS); Model Catalog: STR5029-8-400 (Lot serial: ALL LOTS); Model Catalog: S
  • Product Description
    STERILMED REPROCESSED STRYKER EXTERNAL FIXATION SYSTEMS;STERILMED REPROCESSED DEPUY SYNTHES EXTERNAL FIXATION SYSTEMS;STERILMED REPROCESSED SMITH & NEPHEW EXTERNAL FIXATION SYSTEMS
  • Manufacturer

Manufacturer