Recall of STERI-PROBE RECTAL/ESOPHAGEAL TEMPERATURE PROBE - DISPOSABLE

According to Health Canada (via FOI), this recall involved a device in Canada that was produced by CINCINNATI SUB-ZERO PRODUCTS.

What is this?

A correction or removal action taken by a manufacturer to address a problem with a medical device. Recalls occur when a medical device is defective, when it could be a risk to health, or when it is both defective and a risk to health.

Learn more about the data here
  • Type of Event
    Recall
  • Event ID
    15892
  • Event Risk Class
    II
  • Event Initiated Date
    2011-10-25
  • Event Country
  • Event Source
    HC
  • Notes / Alerts
    Canadian data is current through March 2018. All of the data comes from Health Canada, except for the categories Manufacturer Parent Company and Product Classification.
    The Parent Company and the Product Classification were added by ICIJ.
    The parent company information is based on 2017 public records. The device classification information comes from FDA’s Product Classification by Review Panel, based on matches of recall data from the U.S. and Canada.
  • Extra notes in the data
  • Reason
    Some 491b (p/n 39662) rectal/esophageal temperature probes are not reading correctly causing the blanketrol device which the probes connect to to shut down operation and display a "check probe" error message. when the probe cable the component which connects the probe to the blanketrol device is moved during use the probe intermittently loses continuity causing the probe to inappropriately operate and display/provide wrong temperature values.

Device

  • Model / Serial
    Model Catalog: 491B (Lot serial: >10 numbers contact mfg)
  • Product Description
    Recatal/Esophageal Temperature Probe
  • Manufacturer

Manufacturer

  • Manufacturer Address
    CINCINNATI
  • Manufacturer Parent Company (2017)
  • Source