Recall of STELLARIS PC

According to Health Canada (via FOI), this recall involved a device in Canada that was produced by BAUSCH & LOMB CANADA INC..

What is this?

A correction or removal action taken by a manufacturer to address a problem with a medical device. Recalls occur when a medical device is defective, when it could be a risk to health, or when it is both defective and a risk to health.

Learn more about the data here
  • Type of Event
    Recall
  • Event ID
    110538
  • Event Risk Class
    II
  • Event Initiated Date
    2012-08-31
  • Event Country
  • Event Source
    HC
  • Notes / Alerts
    Canadian data is current through March 2018. All of the data comes from Health Canada, except for the categories Manufacturer Parent Company and Product Classification.
    The Parent Company and the Product Classification were added by ICIJ.
    The parent company information is based on 2017 public records. The device classification information comes from FDA’s Product Classification by Review Panel, based on matches of recall data from the U.S. and Canada.
  • Extra notes in the data
  • Reason
    B-l recently determined that posterior fluidics modules installed in the stellaris pc vision enhancement systems by the b-l field service organization between january 3 2012 and august 8 2012 may need to be recalibrated. these modules which regulate the performance of the vitrectomy probe may exhibit behavior that could result in less efficient aspiration.

Device

  • Model / Serial
    Model Catalog: BL1433 (Lot serial: NA)
  • Product Description
    STELLARIS PC VISION ENHANCEMENT SYSTEM
  • Manufacturer

Manufacturer