Recall of STAR S4 IR EXCIMER LASER SYSTEM

According to Health Canada (via FOI), this recall involved a device in Canada that was produced by AMO CANADA COMPANY.

What is this?

A correction or removal action taken by a manufacturer to address a problem with a medical device. Recalls occur when a medical device is defective, when it could be a risk to health, or when it is both defective and a risk to health.

Learn more about the data here
  • Type of Event
    Recall
  • Event ID
    17126
  • Event Risk Class
    II
  • Event Initiated Date
    2013-09-17
  • Event Country
  • Event Source
    HC
  • Notes / Alerts
    Canadian data is current through March 2018. All of the data comes from Health Canada, except for the categories Manufacturer Parent Company and Product Classification.
    The Parent Company and the Product Classification were added by ICIJ.
    The parent company information is based on 2017 public records. The device classification information comes from FDA’s Product Classification by Review Panel, based on matches of recall data from the U.S. and Canada.
  • Extra notes in the data
  • Reason
    The advisory notice is associated with the visx excimer laser system (licence 2145). this action will address an identified failure of the lamp and illumination system within the laser unit caused by an unexpected failure in the ac power distribution printed circuit board (pcb) in the visx excimer laser system specifically models star s2 star s3 star s4 custom vss and star s4 ir.

Device

  • Model / Serial
    Model Catalog: STAR S4 IR (Lot serial: >10 numbers contact mfg)
  • Product Description
    STAR S4 IR EXCIMER LASER SYSTEM
  • Manufacturer

Manufacturer