Recall of SSP UNITRAY - HIGH RESOLUTION CLASS I

According to Health Canada (via FOI), this recall involved a device in Canada that was produced by LIFE TECHNOLOGIES CORPORATION.

What is this?

A correction or removal action taken by a manufacturer to address a problem with a medical device. Recalls occur when a medical device is defective, when it could be a risk to health, or when it is both defective and a risk to health.

Learn more about the data here
  • Type of Event
    Recall
  • Event ID
    20084
  • Event Risk Class
    III
  • Event Initiated Date
    2010-03-04
  • Event Country
  • Event Source
    HC
  • Notes / Alerts
    Canadian data is current through March 2018. All of the data comes from Health Canada, except for the categories Manufacturer Parent Company and Product Classification.
    The Parent Company and the Product Classification were added by ICIJ.
    The parent company information is based on 2017 public records. The device classification information comes from FDA’s Product Classification by Review Panel, based on matches of recall data from the U.S. and Canada.
  • Extra notes in the data
  • Reason
    This field safety notice serves to notify customers that lane 42 (a24-34d) in the affected a high resolution ssp unitrays does not amplify a*2432 allele. this issue would produce a mistyping for a*2432 positive sample.

Device

  • Model / Serial
    Model Catalog: 47290-10 (Lot serial: 011 536590); Model Catalog: 47290-10 (Lot serial: 011 484916); Model Catalog: 47290-10 (Lot serial: 012 625731)
  • Product Description
    A HIGH RES SSP UNITRAY
  • Manufacturer

Manufacturer