Recall of SERFAS ENERGY SYSTEM - PROBES

According to Health Canada (via FOI), this recall involved a device in Canada that was produced by STRYKER CANADA LP.

What is this?

A correction or removal action taken by a manufacturer to address a problem with a medical device. Recalls occur when a medical device is defective, when it could be a risk to health, or when it is both defective and a risk to health.

Learn more about the data here
  • Type of Event
    Recall
  • Event ID
    88036
  • Event Risk Class
    II
  • Event Initiated Date
    2007-05-01
  • Event Country
  • Event Source
    HC
  • Notes / Alerts
    Canadian data is current through March 2018. All of the data comes from Health Canada, except for the categories Manufacturer Parent Company and Product Classification.
    The Parent Company and the Product Classification were added by ICIJ.
    The parent company information is based on 2017 public records. The device classification information comes from FDA’s Product Classification by Review Panel, based on matches of recall data from the U.S. and Canada.
  • Extra notes in the data
  • Reason
    A possibility of continued activation of the device after the button/foot switch is no longer depressed due to ingress of liquid to the handle from the distal end causing localized burns.

Device

  • Model / Serial
    Model Catalog: 0279-351-100 (Lot serial: 06361AE2 THRU 07044AE2)
  • Product Description
    3.5mm 90-S SERFAS ENERGY SUCTION PROBE
  • Manufacturer

Manufacturer

  • Manufacturer Address
    HAMILTON
  • Manufacturer Parent Company (2017)
  • Source
    HC