Recall of SARNS MODULAR PERFUSION SYSTEM 8000

According to Health Canada (via FOI), this recall involved a device in Canada that was produced by TERUMO CARDIOVASCULAR SYSTEMS CORP..

What is this?

A correction or removal action taken by a manufacturer to address a problem with a medical device. Recalls occur when a medical device is defective, when it could be a risk to health, or when it is both defective and a risk to health.

Learn more about the data here
  • Type of Event
    Recall
  • Event ID
    52855
  • Event Risk Class
    II
  • Event Initiated Date
    2012-08-09
  • Event Country
  • Event Source
    HC
  • Notes / Alerts
    Canadian data is current through March 2018. All of the data comes from Health Canada, except for the categories Manufacturer Parent Company and Product Classification.
    The Parent Company and the Product Classification were added by ICIJ.
    The parent company information is based on 2017 public records. The device classification information comes from FDA’s Product Classification by Review Panel, based on matches of recall data from the U.S. and Canada.
  • Extra notes in the data
  • Reason
    Terumo cardiovascular systems (terumo cvs) has received multiple reports of a malfunction of the cardioplegia or arterial monitor for the sarns modular perfusion system 8000. the malfunctions resulted in intermittent or complete loss of the pressure display on the monitors.

Device

  • Model / Serial
    Model Catalog: 16414 (Lot serial: 1254 (16413)); Model Catalog: 16413 (Lot serial: 2052 (16414)); Model Catalog: 16414 (Lot serial: 1210 (16414)); Model Catalog: 16413 (Lot serial: 1210 (16414)); Model Catalog: 16414 (Lot serial: 1001-3908); Model Catalog: 16413 (Lot serial: 1001-3908); Model Catalog: 16414 (Lot serial: 1001-3840); Model Catalog: 16413 (Lot serial: 1001-3840); Model Catalog: 16414 (Lot serial: 2123 (16414)); Model Catalog: 16413 (Lot serial: 2123 (16414)); Model Catalog: 16414 (Lot serial: 2066 (16414)); Model Catalog: 16413 (Lot serial: 2066 (16414)); Model Catalog: 16414 (Lot serial: 2052 (16414)); Model Catalog: 16413 (Lot serial: 1254 (16413))
  • Product Description
    CARDIOPLEGIA MONITOR AND ARTERIAL MONITOR FOR THE SARNS MODULAR PERFUSION SYSTEM 8000
  • Manufacturer

Manufacturer

  • Manufacturer Address
    ANN ARBOR
  • Manufacturer Parent Company (2017)
  • Source
    HC