Recall of S5 HEART LUNG MACHINE - MAST MOUNTED ROLLER PUMPS

According to Health Canada (via FOI), this recall involved a device in Canada that was produced by LIVANOVA CANADA CORP..

What is this?

A correction or removal action taken by a manufacturer to address a problem with a medical device. Recalls occur when a medical device is defective, when it could be a risk to health, or when it is both defective and a risk to health.

Learn more about the data here
  • Type of Event
    Recall
  • Event ID
    57885
  • Event Risk Class
    II
  • Event Initiated Date
    2012-09-20
  • Event Country
  • Event Source
    HC
  • Notes / Alerts
    Canadian data is current through March 2018. All of the data comes from Health Canada, except for the categories Manufacturer Parent Company and Product Classification.
    The Parent Company and the Product Classification were added by ICIJ.
    The parent company information is based on 2017 public records. The device classification information comes from FDA’s Product Classification by Review Panel, based on matches of recall data from the U.S. and Canada.
  • Extra notes in the data
  • Reason
    The manufacturer recently received reports of a 'fault in motor controller' alarm message. if this occurs the pump stops and audible alarm sounds and a visual alarm is displayed. the 'fault in motor controller' message is displayed on the system panel. on some occasions the circuit breaker of the pump on the emergency power pack may also trip.

Device

  • Model / Serial
    Model Catalog: 10-88-60 (Lot serial: >10 NUMBERS CONTACT MFR); Model Catalog: 10-80-00 (Lot serial: >10 NUMBERS CONTACT MFR); Model Catalog: 10-88-00 (Lot serial: >10 NUMBERS CONTACT MFR)
  • Product Description
    S5 MAST ROLLER PUMP 150 & 85
  • Manufacturer

Manufacturer

  • Manufacturer Address
    MARKHAM
  • Manufacturer Parent Company (2017)
  • Source
    HC