Recall of ROUND MODERATE PLUS PROFILE STERILE RESTERILIZABLE GEL BREAST IMPLANT SIZER

According to Health Canada (via FOI), this recall involved a device in Canada that was produced by JOHNSON & JOHNSON MEDICAL PRODUCTS A DIV. OF JOHNSON & JOHNSON INC..

What is this?

A correction or removal action taken by a manufacturer to address a problem with a medical device. Recalls occur when a medical device is defective, when it could be a risk to health, or when it is both defective and a risk to health.

Learn more about the data here
  • Type of Event
    Recall
  • Event ID
    51994
  • Event Risk Class
    II
  • Event Initiated Date
    2014-05-29
  • Event Country
  • Event Source
    HC
  • Notes / Alerts
    Canadian data is current through March 2018. All of the data comes from Health Canada, except for the categories Manufacturer Parent Company and Product Classification.
    The Parent Company and the Product Classification were added by ICIJ.
    The parent company information is based on 2017 public records. The device classification information comes from FDA’s Product Classification by Review Panel, based on matches of recall data from the U.S. and Canada.
  • Extra notes in the data
  • Reason
    A complaint was received for a volume discrepancy between the pad print (500cc) and laser engraved marking and all other product labeling (550cc) on a mentor gel sizer. during the investigation a second lot was identified as having a style discrepancy between the pad print (moderate profile mp) and all other product labeling (moderate plus m+).

Device

Manufacturer