Recall of ROTAFLOW CENTRIFUGAL PUMP WITH BIOLINE

According to Health Canada (via FOI), this recall involved a device in Canada that was produced by MAQUET-DYNAMED INC..

What is this?

A correction or removal action taken by a manufacturer to address a problem with a medical device. Recalls occur when a medical device is defective, when it could be a risk to health, or when it is both defective and a risk to health.

Learn more about the data here
  • Type of Event
    Recall
  • Event ID
    74016
  • Event Risk Class
    II
  • Event Initiated Date
    2016-03-02
  • Event Country
  • Event Source
    HC
  • Notes / Alerts
    Canadian data is current through March 2018. All of the data comes from Health Canada, except for the categories Manufacturer Parent Company and Product Classification.
    The Parent Company and the Product Classification were added by ICIJ.
    The parent company information is based on 2017 public records. The device classification information comes from FDA’s Product Classification by Review Panel, based on matches of recall data from the U.S. and Canada.
  • Extra notes in the data
  • Reason
    The manufacturer has determined that listed products may possess levels of endotoxin that exceed regulatory limits. mcp has become aware that its sampling plans for endotoxin testing on products labeled "non-pyrogenic" were not compliant with usp requirements for endotoxin on medical devices and ansi/aami st72: 2011 bacterial endotoxins - test methods routine monitoring and alternatives to batch testing. while maquet has not received any complaints of adverse events related to endotoxin exposure maquet cannot ensure devices labeled "non-pyrogenic" are without endotoxin based on a statistically valid sampling plan.

Device

  • Model / Serial
    Model Catalog: BEQ-RF-32 (Lot serial: all lots); Model Catalog: BO-S 2565 (Lot serial: all lots); Model Catalog: BO-S 2564 (Lot serial: all lots); Model Catalog: VKMO 31000 (Lot serial: all lots); Model Catalog: VKMO 30000 (Lot serial: all lots); Model Catalog: BO-VKMO 31000 (Lot serial: all lots); Model Catalog: BO-HQV 104802 (Lot serial: all lots); Model Catalog: BO-HQV 104100 (Lot serial: all lots); Model Catalog: BO-H 104801 (Lot serial: all lots); Model Catalog: BO-H 101111 (Lot serial: all lots); Model Catalog: BO-H65000 (Lot serial: all lots); Model Catalog: RF-32 (Lot serial: all lots); Model Catalog: BEQ-RF-32(USA) (Lot serial: all lots); Model Catalog: BEQ-H 84702 (Lot serial: all lots); Model Catalog: BEQ-H 66607 (Lot serial: all lots); Model Catalog: BEQ-H 66603 (Lot serial: all lots); Model Catalog: BEQ-H 59210 (Lot serial: all lots); Model Catalog: BEQ-H 59209 (Lot serial: all lots); Model Catalog: BEQ-H 56100 (Lot serial: all lots); Model Catalog: BEQ-H 55105 (Lot serial: all lots); Model Cata
  • Product Description
    ROTAFLOW CENTRIFUGAL PUMP WITH BIOLINE;CIRCULATORY CIRCUIT SOFTLINE;QUADROX-I PEDIATRIC OXYGENATOR WITH INTEGRATED HARDSHELL RESERVOIR;PERFUSION SET WITH SOFTLINE COATING;SMALL ADULT HEART LUNG PERFUSION PACK WITH SOFTLINE COATING;ROTAFLOW CENTRIFUGAL PUM