Recall of REJUVENATE MODULAR NECK 8 DEG.

According to Health Canada (via FOI), this recall involved a device in Canada that was produced by STRYKER CANADA LP.

What is this?

A correction or removal action taken by a manufacturer to address a problem with a medical device. Recalls occur when a medical device is defective, when it could be a risk to health, or when it is both defective and a risk to health.

Learn more about the data here
  • Type of Event
    Recall
  • Event ID
    116682
  • Event Risk Class
    II
  • Event Initiated Date
    2012-06-29
  • Event Country
  • Event Source
    HC
  • Notes / Alerts
    Canadian data is current through March 2018. All of the data comes from Health Canada, except for the categories Manufacturer Parent Company and Product Classification.
    The Parent Company and the Product Classification were added by ICIJ.
    The parent company information is based on 2017 public records. The device classification information comes from FDA’s Product Classification by Review Panel, based on matches of recall data from the U.S. and Canada.
  • Extra notes in the data
  • Reason
    Ongoing analysis of the global data following the product correction does not yield a significant increase in the global reported rate for adverse local tissue reaction (altr).However the additional data which includes variability in altr rates among sites may potentially be predictive of an increased likelihood of this condition for both the rejuvenate and abg ii modular hip systems. based on information received to date a product field action to remove these products is being conducted.

Device

  • Model / Serial
    Model Catalog: NLV-340800G (Lot serial: ALL); Model Catalog: NLV-340800Y (Lot serial: ALL); Model Catalog: NLV-380800G (Lot serial: ALL); Model Catalog: NLV-380800Y (Lot serial: ALL); Model Catalog: NLV-420800G (Lot serial: ALL); Model Catalog: NLV-420800Y (Lot serial: ALL); Model Catalog: NLV-300800Y (Lot serial: ALL); Model Catalog: NLV-300800G (Lot serial: ALL); Model Catalog: NLS-421600P (Lot serial: ALL); Model Catalog: NLS-381600P (Lot serial: ALL); Model Catalog: NLS-341600P (Lot serial: ALL); Model Catalog: NLS-301600P (Lot serial: ALL); Model Catalog: SPT-120000S (Lot serial: ALL); Model Catalog: SPT-110000S (Lot serial: ALL); Model Catalog: SPT-100000S (Lot serial: ALL); Model Catalog: SPT-090000S (Lot serial: ALL); Model Catalog: SPT-080000S (Lot serial: ALL); Model Catalog: SPT-070000S (Lot serial: ALL); Model Catalog: NLS-420000B (Lot serial: ALL); Model Catalog: NLS-380000B (Lot serial: ALL); Model Catalog: NLS-340000B (Lot serial: ALL); Model Catalog: NLS-300000B (Lot serial: ALL)
  • Product Description
    REJUVENATE MODULAR NECK 8 DEG.
  • Manufacturer

Manufacturer

  • Manufacturer Address
    HAMILTON
  • Manufacturer Parent Company (2017)
  • Source
    HC