Recall of QUIKHEEL LANCET

According to Health Canada (via FOI), this recall involved a device in Canada that was produced by BECTON DICKINSON CANADA INC..

What is this?

A correction or removal action taken by a manufacturer to address a problem with a medical device. Recalls occur when a medical device is defective, when it could be a risk to health, or when it is both defective and a risk to health.

Learn more about the data here
  • Type of Event
    Recall
  • Event ID
    106892
  • Event Risk Class
    II
  • Event Initiated Date
    2013-07-18
  • Event Country
  • Event Source
    HC
  • Notes / Alerts
    Canadian data is current through March 2018. All of the data comes from Health Canada, except for the categories Manufacturer Parent Company and Product Classification.
    The Parent Company and the Product Classification were added by ICIJ.
    The parent company information is based on 2017 public records. The device classification information comes from FDA’s Product Classification by Review Panel, based on matches of recall data from the U.S. and Canada.
  • Extra notes in the data
  • Reason
    The guide height post (spring pin) dimension of the qhl springs were found damaged and/or the height of the posts were found to be below the lower specification limit. the springs are assembled into the qhl back covers and its main function is to control the activation and retracting blade mechanism. there is a potential that in a small number of cases the lancet will not retract after activation thus creating a potential for exposure to a used sharp.

Device

  • Model / Serial
    Model Catalog: 368101 (Lot serial: 3116007)
  • Product Description
    BD Microtainer Quikheel Lancet
  • Manufacturer

Manufacturer